The significant deviation in the output waveform of a lightning impulse voltage test system can generally be attributed to the following four core reasons:
1.Load Mismatch (Most Common)
Excessive Load Capacitance: Causes the wavefront time to become longer, making the waveform appear "rounded" or "blunt".
Inductive or Resonant Load Structure: Leads to oscillations or overshoot in the waveform.
Load Insulation Failure: Causes the waveform to be chopped (abruptly cut off) before reaching its peak.
2.Generator Circuit Parameter Issues
Incorrect Wavefront Resistance (R_f) or Wave Tail Resistance (R_t) Values: Directly alters the wavefront and wave tail times; an R_f that is too small can also cause oscillations.
Parasitic Inductance and Stray Capacitance: Poor layout or excessively long connecting wires introduce parasitic elements, resulting in glitches or oscillations on the waveform.
3.Measurement System Errors (Crucial)
Poor Divider Response: A voltage divider with inadequate performance (e.g., failing square wave response test) is a primary cause of waveform distortion and overshoot.
Impedance Mismatch: Mismatch between the measurement cable, oscilloscope, and divider impedance causes signal reflections and ringing in the waveform.
4.Grounding and Interference
Poor Grounding: Ground wires not connected at a single point introduce interference, causing baseline drift or noise.
Concise Troubleshooting Steps:
First, perform a no-load test (connect a standard small capacitor, e.g., 500pF). If the waveform is normal, the issue lies with the load.
If the waveform is abnormal even under no-load conditions, the problem is with the generator itself or the measurement system.
Swap in a known-good voltage divider and measurement system for comparison to quickly determine if the issue is measurement-related.